| KVO Rate |
0.1 – 5ml/h (steps : 0.1ml/h) |
| Injection Compliance |
Adaptable for 5 syringe sizes |
| Preset Volume Range |
0.1 ml – 1999.9 ml |
| Occlusion System |
3-level alarming system |
| Mechanical Precision |
≤ ±2 % |
| Injection Rate Range |
| 50 ml Syringe: | 0.1 ml – 999.9 ml/h (step: 0.1 ml/h)<br/>1000 ml – 1800 ml/h (step: 1 ml/h) | 30 ml Syringe: | 0.1 ml – 900 ml/h (step: 0.1 ml/h) | 20 ml Syringe: | 0.1 ml – 600 ml/h (step: 0.1 ml/h) | 10 ml Syringe: | 0.1 ml – 300 ml/h (step: 0.1 ml/h) | 5 ml Syringe: | 0.1 ml – 150 ml/h (step: 0.1 ml/h) |
|
| Purge Rate |
| 50 ml Syringe: | 1800 ml/h (Bolus rate: 1200 ml/h) | 30 ml Syringe: | 900 ml/h (Bolus rate: 600 ml/h) | 20 ml Syringe: | 600 ml/h (Bolus rate: 400 ml/h) | 10 ml Syringe: | 300 ml/h (Bolus rate: 200 ml/h) | 5 ml Syringe: | 150 ml/h (Bolus rate: 100 ml/h) |
|
| Occlusion Alarm Threshold |
| High: | 800 mmHg ± 200 mmHg (106.7 kPa ± 26.7 kPa) | Medium: | 500 mmHg ± 100 mmHg (66.7 kPa ± 13.3 kPa) | Low: | 300 mmHg ± 100 mmHg (40.7 kPa ± 13.3 kPa) |
|
| Fuse |
F1AL/250V (2 nos.) |
| Power Supply |
100 V – 220 V, 50 Hz |
| Battery |
Built–in Rechargeable Li–ion battery (11.1V) Capacity ≥ 2000mAh ; 10 hours of charging Works at 5ml/h after 4 hours of charging |
| Alarm |
Near Injection finished, Preset dose injection finished, Occlusion,
Syringe off, Syringe dislocated, Low Battery, On AC Supply. |
| Safety |
Class I Type BF |
| Operation |
| Ambient Temperature: | 5°C to 40°C | Relative Humidity: | 20% – 90% | Atmospheric Pressure: | 86 kPa – 106 kPa |
|
| Storage |
| Ambient Temperature: | –30°C to 55°C | Relative Humidity: | ≤ 95% | Atmospheric Pressure: | 50 kPa – 106 kPa |
|
| Dimensions |
282 x 210 x 148 mm |
| Weight |
2 Kgs |
- • 4.3” color segment LCD
- •
Multi injection modes – Volume/Time/Body
- •
Adjustable speed, time, volume and dosage
- •
Audio–visual alarms for critical events
Used in general ward, ICU or post-operative care facilities in hospitals to gradually administer intravenous fluid drug dosage in controlled amounts at a predetermined rate and speed through syringe attached with a catheter for consistent and effective infusion therapy.